Pulse-Preamplifiers / Discriminators Series WMT PAD
 
Type   WMT PAD 06
Case   Aluminium
(RF-shielded)
Output to be connect to   Electronic Counter
     
Input    
Amplitude into 50 Ohm mV -3 to > -200
Pulse width (FWHM) ns 5 to 30
Count rate (Poisson distribution) cps up to 2 x 10e6
Impedance Ohm 50, dc-coupled
Threshold mV -3 to -130
Threshold setting by  

2 HEX-switches

in 256 steps

     
Dead Time ns 10-20-30-40-50-60
Selectable   in six steps
Accuracy ns ±2
Drift over specified temperatur
and power suppy range
% ±2
     
Output    
Pulse amplitude into 50 Ohm V

3.2 V open circuit,

1.5 V into 50 Ohm

width ns 50% of dead time
Ratemeter (Count Rate 0...10e6 cps)   0...-4V, accuracy 5%
     
Power supply V +6, -6
Accuracy % ±5
Connector   9-pin D-Sub
     
Temperatur Range    
For specified data °C 20 to 40
Operating °C 0 to 50
     
* depends on pulse amplitude and threshold setting
     
     
return